Precision of flash pyrolysis-GC-MS analysis: Experimental evaluation of the sources of variability
Abstract
The present study was carried out to evaluate the precision and to identify significant sources of variability in flash pyrolysis-GC-MS experiments. The analysis of variance has been applied to the example of polyethylene pyrolysis to estimate the impact of several controlled factors: sample handling, flash pyrolysis filament ageing and calibration. It was demonstrated that to get the best precision in experimental results, it is recommended to set the operator to reduce the impact of sample handling procedure and the filament to avoid variability due to changes in filament characteristics. In comparison, filament ageing and filament calibration are shown to influence in a limited extent the experimental results. Cop. 2009 Vieweg+Teubner / GWV Fachverlage GmbH.